OwenDuffy.net 


Two point thermistor calibration

This calculator calculates R25 and B for the thermistor model \( T=\frac1{\frac1{T_0}+\frac1{B} ln \frac{R}{R_0}}\) where T and To are absolute (K).

Use this calculator to determine R25 and B from T,R pairs measured from either a thermistor or thermometer.

Inputs:    
  T (°C) R (Ω)
Measurements
Results:
R25  
B  

The model is an approximate one, and you will likely get different parameters for different measurement points... though hopefully only a small change. The best thing to do is to use data points towards each end of the intended operating range. It also helps if the first data point is close to 25°.

References / Links

Changes

Version Date Description
1.01 12/06/2015 Initial.
1.02    
1.03    
1.04    
1.05    